Alpha-Beta radiation detector performs the following functions: measurement of α- β- radiation flux density, search for α- β- radiation sources.
BDAB can be used by radiological and isotope tracing laboratories and customs officers, first responders, police, emergency teams.
within range from 1.0 to 5∙105 min-1·cm-2 | ± (20 + А/φ) %, where А – coefficient equal to 10 min-1·cm-2; φ – measured density of α- radiation flux |
within range from 10,0 to 106 min-1·cm-2 | ± (20 + А/φ) %, where А – coefficient equal to 100 min-1·cm-2; φ – measured density of α- radiation flux |
Operation Principle
A built-in detector based on the proportional counter is used to register a- and β- radiation flux density (DER, count rate). Data communication channel enables information exchange with Spectroscopic Radiation Monitor РM1403 / Display and Control Unit PM1403 by RS-485 interface.
BDAB enables the user to set and constantly monitor the threshold level of α- and β- radiation fluxes through the whole measurement range and record the set threshold level of α- and β- radiation density values in the instrument’s non-volatile memory.
Spectroscopic Radiation Monitor PM1403 Operating Mode
When the instrument is connected to the Spectroscopic Radiation Monitor РМ1403 / Display and Control Unit РМ1403 / Display and Control Unit РМ1403-01 or PC it enables:
- α- measurement of β- radiation flux density
- α- measurement of β- radiation flux density
- search for α- and β- radiation sources
- program operation modes
Radiation Detectors Set PM1403
BDG2 - GM detector, gamma radiation detector
BDG3 - Csl(Ti) detector, gamma radiation detector
BDN - He-3 detector, neutron radiation detector
BDAB - proportional counter, alpha-beta radiation detector
Detector
|
proportional counter | ||||
Energy range
|
15 KeV – 3.5 MeV (β- particles) | ||||
α-radiation flux density measurement range
|
from 1.0 to 5∙105 min-1cm-2 | ||||
β-radiation flux density measurement range
|
from 10.0 to 106 min-1·cm-2 | ||||
α-radiation flux density accuracy
|
|
||||
β-radiation flux density accuracy
|
|
||||
Instrument sensitivity
|
|
||||
Data transfer
|
RS-485 or USB interface | ||||
Dimensions
|
71х45х130 mm | ||||
Weight
|
0.45 kg | ||||
Ingress protection
|
IP64 | ||||
Drop test on concrete floor
|
0.7m | ||||
Mean recovery time
|
60 min | ||||
Mean service lifetime
|
10 years | ||||
Atmospheric pressure
|
from 84 to 106.7 kPa | ||||
Operating temperature
|
from -20 to +50 °С | ||||
Humidity
|
up to 95 % at 35 °С |