Gamma-Neutron PRDs
On order
Gamma-neutron personal radiation detectors PM1401GNM are designed to detect the slightest amounts of gamma and neutron radioactive materials in harsh environmental conditions and difficult-to-access areas. Instruments are lightweight, have hermetic and shockproof case, non-volatile memory for data storage, small size and user-friendly interface.
Detector
—
CsI(Tl) scintillator, 3He, Geiger-Muller tube
Energy range of the measures of neutron radiation
—
0.025 eV – 14 MeV
Energy range of the measures of photon radiation
—
33 keV – 3 MeV
Dose equivalent rate (DER) measurement range
—
0.1 – 9999 μSv/h |
DER indication range
—
0.01 – 9999 μSv/h |
Neutron count rate indication range
—
1.0 - 99 с-1 |
DER measurement accuracy
—
±20% |
Sensitivity to neutron radiation
—
for Pu-α-Be | 0.1 (counts·cm2)/neutron |
for heat neutrons | 7.0 (counts·cm2)/neutron |
for Pu-a-Be, at use of chamber-moderator or on phantom | 1.0 (counts·cm2)/neutron |
Time-to-alarn; neutrons
—
2 s (corresponds to ANSI N42.32)
Time-to-alarm; photons
—
2 s (corresponds to ANSI N42.32)
Alarm Type
—
visual, audio, vibration (external)
PC communication
—
USB
Dimensions
—
195х57х32 mm
Weight
—
0.45 kg
Ingress protection
—
IP65
Drop test on concrete floor
—
0.7 m
Atmospheric pressure
—
from 84 to 106.7 kPa
Operating temperature
—
-30°C to 50°C; (LCD from -15°C to +50 °C)
Humidity
—
up to 98 % at 35 °C (+95 °F)
Supply voltage
—
1.5 (+ 0.1; - 0.4) V (one АА battery (LR6))
Battery Lifetime
—
no less 800 h